General Handpiece Maintenance & Cleaning Guide

Maintenance for after use and before Autoclaving

Step: 1 Preparation

  • Detach handpiece from tubing.
  • Remove bur from Handpiece.

DO NOT submerge handpiece in any solution.

MC Repairs

Step:2 Clean

  • Wipe away any external debris using a non-alcoholic approved disinfectant wipe.
  • Using a dry tissue, remove any excess detergent leftover on the handpiece from the wipe.

DO NOT use any chemical solution or submerge in water.

MC Repairs

Step: 3 Connect

  • Select the correct oil adaptor for the handpiece you want to lubricate.
  • Ensure the nozzle is securely attached.
  • Shake the spray can well before attaching to the handpiece.
MC Repairs

Step:4 Spray

  • Insert the oil adaptor into the handpiece.
  • Always hold the spray can upright.
  • Spray lubrication oil through the handpiece 2-3 seconds onto a clean tissue.

Please note: If the oil from the handpiece head has a brown discoloration, spray two more times until the oil runs clear.

MC Repairs

Step:5 Check

  • Repeat steps 3 and 4 until the oil runs clear.
  • If the oil remains discolored after three sprays, send the handpiece for servicing, as this may indicate a rust issue.
  • Package in a sterilised pouch accordance with appplicable standards.

If the oil comes out of the handpiece head with black discoloration, complete the sterilisation process, then send the handpiece for service before using it again.

MC Repairs

Step:6 Sterilise

  • Before placing your equipment in the autoclave, ensure you’ve selected the correct cycle and verified that the autoclave temperature does not exceed 138°C .
  • Load the items into the autoclave chamber according to the manufacturer’s guidelines.
  • If you’re using a non-vacuum autoclave, you’ll need to repeat step 4 following cycle completion.

For any further advice or guidance, please contact us.

MC Repairs

Best Practices for cleaning and maintaining Equipment:

Avoid Chemical Disinfectants: Refrain from using chemical disinfectants, abrasive cleaners, or antibacterial soaps as cleaners

Avoid Submerging: Avoid submerging any handpieces into any type of liquid for cleaning or disinfection.

Avoid Ultrasonic Cleaners: Do not utilise ultrasonic cleaners for cleaning any handpieces.

Ensure Proper Installation: Ensure all instruments are removed from your handpiece or equipment prior to cleaning and lubricating

Autoclave: When autoclaving do not exceed a temperature of 138°C and do not overfill trays as this will prevent full sterilisation.

Please note: These are generalised guidelines as recommended by MC Repairs Ltd. For exact guidelines, please refer to Lubrication System user manual.

Maintenance Support

Your Printable Guides

You can print or request our maintenance guides, a handy printout to use for reference of essential day-to-day care of some of your key dental equipment. Fill out our contact form with your details to request one today.

MC Repairs
Download

Your Maintenance Guide

Send Your Repair – It’s Quick & Easy!

Getting your dental equipment repaired has never been simpler!
Choose the option that suits you best:

Label
Option 1

Print Your Own (Fast & Simple!)

  • Download & print your repair form and FREE tracked returns label.
  • Pack your repair securely and drop it 
off at any Post Office.
  • No upfront shipping costs – 
we handle it!
Repair Pack
Option 2

Let Us Collect It (Hassle-Free!)

  • No printer? No problem! We’ll arrange a courier collection for you.
  • Our approved courier will pick up your repair from your chosen address.
  • Ideal for larger or multiple items.
Option 3

Request a Repair Pack (We Send Everything!)

  • Not sure how to pack your repair? We’ll send you a pre-labelled repair kit.
  • Includes protective packaging & return instructions.
  • The easiest way to ensure safe transit.

Why Choose MC Repairs?

Fully Accredited

Fully Accredited

We’re fully accredited and equipped for all small dental equipment repairs on brands such as Acteon, Anthogyr, Bien Air, DTE, EMS, KaVo, Mk-Dent, NSK, Star Dental, W&H, Woodpecker and much more!

OEM Quality Components

OEM Quality Components

Where possible, we offer our customers only the best quality components, including longer warranties and ease of mind in manufacturers own parts so your equipment is as good as new!

State-of-the-Art Workshop

State-of-the-Art Workshop

We don’t just say it, we actually mean it!
We’ve invested heavily into building our new state-of-the art workshop, adding much needed precision tooling and testing equipment.

Sustainable Repairs

Sustainable Repairs

Our dedication to environmentally responsible repairs has earned us ISO 14001 certification. Ensuring that our repair processes are fully efficient and eco-friendly.

Excellent Customer Service

Excellent Customer Service

Whether you order online or over the phone, our team are here to resolve any queries and deliver beyond dental practice’s expectations.

No Obligation, FREE Quotation

No Obligation, FREE Quotation

We adhere to our No Obligation, FREE Quotation allowing for in-depth analysis without any condition to pay unless you’re satisfied with the costs!

Can't Find Your Product?

If you don’t see your product, call us for more information or use our contact form here.

MC Repairs
Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

For a detailed breakdown of the cookies we use and more information please see our cookie policy page.